Thursday, June 19, 2008

What are the Artifacts of CT?


Although CT is a relatively accurate test, it is liable to produce artifacts, such as the following.

Example of Beam Hardening
Aliasing Artifact or Streaks
These appear as dark lines which radiate away from sharp corners. It occurs because it is impossible for the scanner to 'sample' or take enough projections of the object, which is usually metallic. It can also occur when an insufficient X-ray tube current is selected, and insufficient penetration of the x-ray occurs. These artifacts are also closely tied to motion during a scan. This type of artifact commonly occurs in head images around the pituitary fossa area.
Partial Volume Effect
This appears as 'blurring' over sharp edges. It is due to the scanner being unable to differentiate between a small amount of high-density material (e.g. bone) and a larger amount of lower density (e.g. cartilage). The processor tries to average out the two densities or structures, and information is lost. This can be partially overcome by scanning using thinner slices.
Ring Artifact
Probably the most common mechanical artifact, the image of one or many 'rings' appears within an image. This is usually due to a detector fault.
Noise Artifact
This appears as graining on the image and is caused by a low signal to noise ratio. This occurs more commonly when a thin slice thickness is used. It can also occur when the kV or mA of the X-ray tube is insufficient to penetrate the anatomy.
Motion Artifact
This is seen as blurring and/or streaking which is caused by movement of the object being imaged.
Windmill
Streaking appearances can occur when the detectors intersect the reconstruction plane. This can be reduced with filters or a reduction in pitch.
Beam Hardening
This can give a 'cupped appearance'. It occurs when there is more attenuation in the center of the object than around the edge. This is easily corrected by filtration and software.

No comments: